Technical Specification TSFLUXUS_F72xV3-0-0-0EN_Leu

Flexim FLUXUS F721, F722 Ultrasonic Flowmeter

![image 1](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile1.png)

![image 2](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile2.png)

Permanently Installed Ultrasonic Flowmeter for Liquids Features

  • • Exact and highly reliable clamp-on volume and mass flow measurement
  • • High measurement accuracy even at very low as well as very high flow rates and independent of the flow direction (bidirectional)
  • • The measurement is zero point stable, drift free and independent of pipe material, process pressure, process temperature and process fluid
  • ##### Applications

    • Chemical industry, petrochemical industry, oil and gas industry, pharmaceutical industry, semiconductor industry, manufacturing industries, building technology/energy management, water and wastewater industry, mining industries

    ![image 3](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile3.png)

    ![image 4](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile4.png)

    Transmitter Technical data

    | | |FLUXUS F721-NNN-*A F721-NNN-*S|FLUXUS F721-A2N-*A F721-A2N-*S|FLUXUS F721-F2N-*A F721-F2N-*S|FLUXUS F722-NNN-*A F722-NNN-*S|FLUXUS F722-A2N-*A F722-A2N-*S|FLUXUS F722-F2N-*A F722-F2N-*S| |---|---|---|---|---|---|---|---| | | |![image 5](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile5.png)

    ![image 6](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile6.png)|![image 5](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile5.png)

    ![image 6](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile6.png)|![image 5](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile5.png)

    ![image 6](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile6.png)|![image 5](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile5.png)

    ![image 6](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile6.png)|![image 5](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile5.png)

    ![image 6](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile6.png)|![image 5](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile5.png)

    ![image 6](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile6.png)| |design| |standard field device|standard field device zone 2|standard field device FM Class I Div. 2|standard field device|standard field device zone 2|standard field device FM Class I Div. 2| |measurement|measurement|measurement|measurement|measurement|measurement|measurement|measurement| |measurement principle| |transit time difference correlation principle, automatic NoiseTrek selection for measurements with high gaseous or solid content|transit time difference correlation principle, automatic NoiseTrek selection for measurements with high gaseous or solid content|transit time difference correlation principle, automatic NoiseTrek selection for measurements with high gaseous or solid content|transit time difference correlation principle, automatic NoiseTrek selection for measurements with high gaseous or solid content|transit time difference correlation principle, automatic NoiseTrek selection for measurements with high gaseous or solid content|transit time difference correlation principle, automatic NoiseTrek selection for measurements with high gaseous or solid content| |synchronised channel averaging| |-|-|-|x (2 measuring channels necessary)|x (2 measuring channels necessary)|x (2 measuring channels necessary)| |flow velocity|m/s|0.01...25|0.01...25|0.01...25|0.01...25|0.01...25|0.01...25| |repeatability| |0.15 % MV ±0.005 m/s|0.15 % MV ±0.005 m/s|0.15 % MV ±0.005 m/s|0.15 % MV ±0.005 m/s|0.15 % MV ±0.005 m/s|0.15 % MV ±0.005 m/s| |fluid| |all acoustically conductive liquids with < 10 % gaseous or solid content in volume (transit time difference principle)|all acoustically conductive liquids with < 10 % gaseous or solid content in volume (transit time difference principle)|all acoustically conductive liquids with < 10 % gaseous or solid content in volume (transit time difference principle)|all acoustically conductive liquids with < 10 % gaseous or solid content in volume (transit time difference principle)|all acoustically conductive liquids with < 10 % gaseous or solid content in volume (transit time difference principle)|all acoustically conductive liquids with < 10 % gaseous or solid content in volume (transit time difference principle)| |temperature compensation| |corresponding to the recommendations in ANSI/ASME MFC-5.1-2011|corresponding to the recommendations in ANSI/ASME MFC-5.1-2011|corresponding to the recommendations in ANSI/ASME MFC-5.1-2011|corresponding to the recommendations in ANSI/ASME MFC-5.1-2011|corresponding to the recommendations in ANSI/ASME MFC-5.1-2011|corresponding to the recommendations in ANSI/ASME MFC-5.1-2011| |measurement uncertainty (volumetric flow rate)|measurement uncertainty (volumetric flow rate)|measurement uncertainty (volumetric flow rate)|measurement uncertainty (volumetric flow rate)|measurement uncertainty (volumetric flow rate)|measurement uncertainty (volumetric flow rate)|measurement uncertainty (volumetric flow rate)|measurement uncertainty (volumetric flow rate)| |measurement uncertainty of the measuring system1| |±0.3 % MV ±0.005 m/s|±0.3 % MV ±0.005 m/s|±0.3 % MV ±0.005 m/s|±0.3 % MV ±0.005 m/s|±0.3 % MV ±0.005 m/s|±0.3 % MV ±0.005 m/s| |measurement uncertainty at the measuring point2| |±1 % MV ±0.005 m/s|±1 % MV ±0.005 m/s|±1 % MV ±0.005 m/s|±1 % MV ±0.005 m/s|±1 % MV ±0.005 m/s|±1 % MV ±0.005 m/s|

    Page 2

    |transmitter|transmitter|transmitter|transmitter|transmitter|transmitter|transmitter|transmitter| |power supply| |• 100...230 V/50...60 Hz or
    • 20...32 V DC or
    • 11...16 V DC
    |• 100...230 V/50...60 Hz or
    • 20...32 V DC or
    • 11...16 V DC
    |• 100...230 V/50...60 Hz or
    • 20...32 V DC or
    • 11...16 V DC
    |• 100...230 V/50...60 Hz or
    • 20...32 V DC or
    • 11...16 V DC
    |• 100...230 V/50...60 Hz or
    • 20...32 V DC or
    • 11...16 V DC
    |• 100...230 V/50...60 Hz or
    • 20...32 V DC or
    • 11...16 V DC
    | |power consumption|W|< 15|< 15|< 15|< 15|< 15|< 15| |number of measuring channels| |1, optional: 2|1, optional: 2|1, optional: 2|1, optional: 2 (1 measuring point)|1, optional: 2 (1 measuring point)|1, optional: 2 (1 measuring point)| |damping|s|0...100 (adjustable)|0...100 (adjustable)|0...100 (adjustable)|0...100 (adjustable)|0...100 (adjustable)|0...100 (adjustable)| |measuring cycle|Hz|100...1000 (1 channel)|100...1000 (1 channel)|100...1000 (1 channel)|100...1000 (1 channel)|100...1000 (1 channel)|100...1000 (1 channel)| |response time|s|1 (1 channel), option: 0.02|1 (1 channel), option: 0.02|1 (1 channel), option: 0.02|1 (1 channel), option: 0.02|1 (1 channel), option: 0.02|1 (1 channel), option: 0.02| |housing material| |aluminum, powder coated or stainless steel 316L (1.4404)|aluminum, powder coated or stainless steel 316L (1.4404)|aluminum, powder coated or stainless steel 316L (1.4404)|aluminum, powder coated or stainless steel 316L (1.4404)|aluminum, powder coated or stainless steel 316L (1.4404)|aluminum, powder coated or stainless steel 316L (1.4404)| |degree of protection| |IP66|IP66|aluminum housing: IP66/NEMA 4X stainless steel housing: IP65|IP66|IP66|aluminum housing: IP66/NEMA 4X stainless steel housing: IP65| |dimensions|mm|see dimensional drawing|see dimensional drawing|see dimensional drawing|see dimensional drawing|see dimensional drawing|see dimensional drawing| |weight|kg|aluminum housing: 5.4 stainless steel housing: 5.1|aluminum housing: 5.4 stainless steel housing: 5.1|aluminum housing: 5.4 stainless steel housing: 5.1|aluminum housing: 5.4 stainless steel housing: 5.1|aluminum housing: 5.4 stainless steel housing: 5.1|aluminum housing: 5.4 stainless steel housing: 5.1| |fixation| |wall mounting, optional: 2" pipe mounting|wall mounting, optional: 2" pipe mounting|wall mounting, optional: 2" pipe mounting|wall mounting, optional: 2" pipe mounting|wall mounting, optional: 2" pipe mounting|wall mounting, optional: 2" pipe mounting| |ambient temperature|°C|-40...+60 (< -20 without operation of the display)|-40...+60 (< -20 without operation of the display)|aluminum housing: 40...+55/60 (< -20 without operation of the display) stainless steel housing: -20...+55/ 60|-40...+60 (< -20 without operation of the display)|-40...+60 (< -20 without operation of the display)|aluminum housing: 40...+55/60 (< -20 without operation of the display) stainless steel housing: -20...+55/ 60| |display| |128 x 64 pixels, backlight|128 x 64 pixels, backlight|128 x 64 pixels, backlight|128 x 64 pixels, backlight|128 x 64 pixels, backlight|128 x 64 pixels, backlight| |menu language| |English, German, French, Spanish, Dutch, Russian, Polish, Turkish, Italian|English, German, French, Spanish, Dutch, Russian, Polish, Turkish, Italian|English, German, French, Spanish, Dutch, Russian, Polish, Turkish, Italian|English, German, French, Spanish, Dutch, Russian, Polish, Turkish, Italian|English, German, French, Spanish, Dutch, Russian, Polish, Turkish, Italian|English, German, French, Spanish, Dutch, Russian, Polish, Turkish, Italian| |explosion protection|explosion protection|explosion protection|explosion protection|explosion protection|explosion protection|explosion protection|explosion protection| |• ATEX/IECEx|• ATEX/IECEx|• ATEX/IECEx|• ATEX/IECEx|• ATEX/IECEx|• ATEX/IECEx|• ATEX/IECEx|• ATEX/IECEx| |marking| |-|F721-A20*A, F721-A20*S:

    Ex nA nC ic IIC T4 Gc Ex tb IIIC T120 °C Db Ta -40...+60 °C

    II3G II2D

    0637

    |-|-|F722-A20*A, F722-A20*S:

    Ex nA nC ic IIC T4 Gc Ex tb IIIC T120 °C Db Ta -40...+60 °C

    II3G II2D

    0637

    |-| |certification| |-|IBExU11ATEX1015, IECEx IBE 11.0008|-|-|IBExU11ATEX1015, IECEx IBE 11.0008|-|

  • 1 with aperture calibration of the transducers
  • 2 for transit time difference principle and reference conditions
  • 3 outside the explosive atmosphere (housing cover open)
  • | | |FLUXUS F721-NNN-*A F721-NNN-*S|FLUXUS F721-A2N-*A F721-A2N-*S|FLUXUS F721-F2N-*A F721-F2N-*S|FLUXUS F722-NNN-*A F722-NNN-*S|FLUXUS F722-A2N-*A F722-A2N-*S|FLUXUS F722-F2N-*A F722-F2N-*S| |---|---|---|---|---|---|---|---| |• FM|• FM|• FM|• FM|• FM|• FM|• FM|• FM| |marking| |-|-|F721-F202, F721-F203:

    NI/Cl. I,II,III/ Div. 2/GP. A,B,C,D,E, F,G/

    T5 F721-F201:

    NI/Cl. I,II,III/ Div. 2/GP. A,B,C,D,E, F,G/ T4A

    |-|-|F722-F202, F722-F203:

    NI/Cl. I,II,III/ Div. 2/GP. A,B,C,D,E, F,G/

    T5 F722-F201:

    NI/Cl. I,II,III/ Div. 2/GP. A,B,C,D,E, F,G/ T4A

    | |measuring functions|measuring functions|measuring functions|measuring functions|measuring functions|measuring functions|measuring functions|measuring functions| |physical quantities| |volumetric flow rate, mass flow rate, flow velocity, thermal energy rate (if temperature inputs are installed)|volumetric flow rate, mass flow rate, flow velocity, thermal energy rate (if temperature inputs are installed)|volumetric flow rate, mass flow rate, flow velocity, thermal energy rate (if temperature inputs are installed)|volumetric flow rate, mass flow rate, flow velocity, thermal energy rate (if temperature inputs are installed)|volumetric flow rate, mass flow rate, flow velocity, thermal energy rate (if temperature inputs are installed)|volumetric flow rate, mass flow rate, flow velocity, thermal energy rate (if temperature inputs are installed)| |totaliser| |volume, mass, optional: thermal energy|volume, mass, optional: thermal energy|volume, mass, optional: thermal energy|volume, mass, optional: thermal energy|volume, mass, optional: thermal energy|volume, mass, optional: thermal energy| |calculation functions| |average, difference, sum (2 measuring channels necessary)|average, difference, sum (2 measuring channels necessary)|average, difference, sum (2 measuring channels necessary)|average, difference, sum (2 measuring channels necessary)|average, difference, sum (2 measuring channels necessary)|average, difference, sum (2 measuring channels necessary)| |diagnostic functions| |sound speed, signal amplitude, SNR, SCNR, standard deviation of amplitudes and transit times|sound speed, signal amplitude, SNR, SCNR, standard deviation of amplitudes and transit times|sound speed, signal amplitude, SNR, SCNR, standard deviation of amplitudes and transit times|sound speed, signal amplitude, SNR, SCNR, standard deviation of amplitudes and transit times|sound speed, signal amplitude, SNR, SCNR, standard deviation of amplitudes and transit times|sound speed, signal amplitude, SNR, SCNR, standard deviation of amplitudes and transit times| |communication interfaces|communication interfaces|communication interfaces|communication interfaces|communication interfaces|communication interfaces|communication interfaces|communication interfaces| |service interfaces| |measured value transmission, parametrisation of the transmitter:

    • USB3
    • LAN3
    |measured value transmission, parametrisation of the transmitter:

    • USB3
    • LAN3
    |measured value transmission, parametrisation of the transmitter:

    • USB3
    • LAN3
    |measured value transmission, parametrisation of the transmitter:

    • USB3
    • LAN3
    |measured value transmission, parametrisation of the transmitter:

    • USB3
    • LAN3
    |measured value transmission, parametrisation of the transmitter:

    • USB3
    • LAN3
    | |process interfaces| |max. 1 option:

    • RS485 (ASCII sender)
    • Modbus RTU
    • BACnet MS/TP
    • M-Bus
    • HART
    • Profibus PA
    • FF H1
    • Modbus TCP
    • BACnet IP
    |max. 1 option:

    • RS485 (ASCII sender)
    • Modbus RTU
    • BACnet MS/TP
    • HART
    • Profibus PA
    • FF H1
    • Modbus TCP
    • BACnet IP
    |max. 1 option:

    • RS485 (ASCII sender)
    • Modbus RTU
    • BACnet MS/TP
    • HART
    • Profibus PA
    • FF H1
    • Modbus TCP
    • BACnet IP
    |max. 1 option:

    • RS485 (ASCII sender)
    • Modbus RTU
    • BACnet MS/TP
    • M-Bus
    • HART
    • Profibus PA
    • FF H1
    • Modbus TCP
    • BACnet IP
    |max. 1 option:

    • RS485 (ASCII sender)
    • Modbus RTU
    • BACnet MS/TP
    • HART
    • Profibus PA
    • FF H1
    • Modbus TCP
    • BACnet IP
    |max. 1 option:

    • RS485 (ASCII sender)
    • Modbus RTU
    • BACnet MS/TP
    • HART
    • Profibus PA
    • FF H1
    • Modbus TCP
    • BACnet IP
    | |accessories|accessories|accessories|accessories|accessories|accessories|accessories|accessories|

    Page 3

    |data transmission kit| |USB cable|USB cable|USB cable|USB cable|USB cable|USB cable| |software| |• FluxDiagReader: reading of measured values and parameters, graphical representation
    • FluxDiag (optional): reading of measurement data, graphical representation, report generation, parametrisation of the transmitter
    |• FluxDiagReader: reading of measured values and parameters, graphical representation
    • FluxDiag (optional): reading of measurement data, graphical representation, report generation, parametrisation of the transmitter
    |• FluxDiagReader: reading of measured values and parameters, graphical representation
    • FluxDiag (optional): reading of measurement data, graphical representation, report generation, parametrisation of the transmitter
    |• FluxDiagReader: reading of measured values and parameters, graphical representation
    • FluxDiag (optional): reading of measurement data, graphical representation, report generation, parametrisation of the transmitter
    |• FluxDiagReader: reading of measured values and parameters, graphical representation
    • FluxDiag (optional): reading of measurement data, graphical representation, report generation, parametrisation of the transmitter
    |• FluxDiagReader: reading of measured values and parameters, graphical representation
    • FluxDiag (optional): reading of measurement data, graphical representation, report generation, parametrisation of the transmitter
    | |data logger|data logger|data logger|data logger|data logger|data logger|data logger|data logger| |loggable values| |all physical quantities, totalised physical quantities and diagnostic values|all physical quantities, totalised physical quantities and diagnostic values|all physical quantities, totalised physical quantities and diagnostic values|all physical quantities, totalised physical quantities and diagnostic values|all physical quantities, totalised physical quantities and diagnostic values|all physical quantities, totalised physical quantities and diagnostic values| |capacity| |max. 800 000 measured values|max. 800 000 measured values|max. 800 000 measured values|max. 800 000 measured values|max. 800 000 measured values|max. 800 000 measured values| |outputs|outputs|outputs|outputs|outputs|outputs|outputs|outputs| | | |The outputs are galvanically isolated from the transmitter.|The outputs are galvanically isolated from the transmitter.|The outputs are galvanically isolated from the transmitter.|The outputs are galvanically isolated from the transmitter.|The outputs are galvanically isolated from the transmitter.|The outputs are galvanically isolated from the transmitter.| |number| |on request|on request|on request|on request|on request|on request| |• switchable current output|• switchable current output|• switchable current output|• switchable current output|• switchable current output|• switchable current output|• switchable current output|• switchable current output| | | |All switchable current outputs are jointly switched to active or passive.|All switchable current outputs are jointly switched to active or passive.|All switchable current outputs are jointly switched to active or passive.|All switchable current outputs are jointly switched to active or passive.|All switchable current outputs are jointly switched to active or passive.|All switchable current outputs are jointly switched to active or passive.| |range|mA|4...20 (3.2...22)|4...20 (3.2...22)|4...20 (3.2...22)|4...20 (3.2...22)|4...20 (3.2...22)|4...20 (3.2...22)| |accuracy| |0.04 % MV ±3 μA|0.04 % MV ±3 μA|0.04 % MV ±3 μA|0.04 % MV ±3 μA|0.04 % MV ±3 μA|0.04 % MV ±3 μA| |active output| |Rext < 250 Ω|Rext < 250 Ω|Rext < 250 Ω|Rext < 250 Ω|Rext < 250 Ω|Rext < 250 Ω| |passive output| |Uext = 8...30 V, depending on Rext (Rext < 1 kΩ at 30 V)|Uext = 8...30 V, depending on Rext (Rext < 1 kΩ at 30 V)|Uext = 8...30 V, depending on Rext (Rext < 1 kΩ at 30 V)|Uext = 8...30 V, depending on Rext (Rext < 1 kΩ at 30 V)|Uext = 8...30 V, depending on Rext (Rext < 1 kΩ at 30 V)|Uext = 8...30 V, depending on Rext (Rext < 1 kΩ at 30 V)| |• HART|• HART|• HART|• HART|• HART|• HART|• HART|• HART| |range|mA|4...20|4...20|4...20|4...20|4...20|4...20| |accuracy| |0.1 % MV ±15 μA|0.1 % MV ±15 μA|0.1 % MV ±15 μA|0.1 % MV ±15 μA|0.1 % MV ±15 μA|0.1 % MV ±15 μA| |active output| |Uint = 24 V, Rext < 500 Ω|Uint = 24 V, Rext < 500 Ω|Uint = 24 V, Rext < 500 Ω|Uint = 24 V, Rext < 500 Ω|Uint = 24 V, Rext < 500 Ω|Uint = 24 V, Rext < 500 Ω| |passive output| |Uext = 10...24 V DC, depending on Rext (Rext < 1 kΩ at 24 V)|Uext = 10...24 V DC, depending on Rext (Rext < 1 kΩ at 24 V)|Uext = 10...24 V DC, depending on Rext (Rext < 1 kΩ at 24 V)|Uext = 10...24 V DC, depending on Rext (Rext < 1 kΩ at 24 V)|Uext = 10...24 V DC, depending on Rext (Rext < 1 kΩ at 24 V)|Uext = 10...24 V DC, depending on Rext (Rext < 1 kΩ at 24 V)| |• voltage output|• voltage output|• voltage output|• voltage output|• voltage output|• voltage output|• voltage output|• voltage output| |range|V|0...1 or 0...10|0...1 or 0...10|0...1 or 0...10|0...1 or 0...10|0...1 or 0...10|0...1 or 0...10| |accuracy| |0...1 V: 0.1 % MV ±1 mV 0...10 V: 0.1 % MV ±10 mV|0...1 V: 0.1 % MV ±1 mV 0...10 V: 0.1 % MV ±10 mV|0...1 V: 0.1 % MV ±1 mV 0...10 V: 0.1 % MV ±10 mV|0...1 V: 0.1 % MV ±1 mV 0...10 V: 0.1 % MV ±10 mV|0...1 V: 0.1 % MV ±1 mV 0...10 V: 0.1 % MV ±10 mV|0...1 V: 0.1 % MV ±1 mV 0...10 V: 0.1 % MV ±10 mV| |internal resistance| |Rint = 500 Ω|Rint = 500 Ω|Rint = 500 Ω|Rint = 500 Ω|Rint = 500 Ω|Rint = 500 Ω| |• frequency output|• frequency output|• frequency output|• frequency output|• frequency output|• frequency output|• frequency output|• frequency output| |range|kHz|0...5|0...5|0...5|0...5|0...5|0...5| |optorelay| |24 V/4 mA, Rint = 66.5 Ω|24 V/4 mA, Rint = 66.5 Ω|24 V/4 mA, Rint = 66.5 Ω|24 V/4 mA, Rint = 66.5 Ω|24 V/4 mA, Rint = 66.5 Ω|24 V/4 mA, Rint = 66.5 Ω|

  • 1 with aperture calibration of the transducers
  • 2 for transit time difference principle and reference conditions
  • 3 outside the explosive atmosphere (housing cover open)
  • | | |FLUXUS F721-NNN-*A F721-NNN-*S|FLUXUS F721-A2N-*A F721-A2N-*S|FLUXUS F721-F2N-*A F721-F2N-*S|FLUXUS F722-NNN-*A F722-NNN-*S|FLUXUS F722-A2N-*A F722-A2N-*S|FLUXUS F722-F2N-*A F722-F2N-*S| |---|---|---|---|---|---|---|---| |• digital output|• digital output|• digital output|• digital output|• digital output|• digital output|• digital output|• digital output| |functions| |• frequency output
    • binary output
    • pulse output
    |• frequency output
    • binary output
    • pulse output
    |• frequency output
    • binary output
    • pulse output
    |• frequency output
    • binary output
    • pulse output
    |• frequency output
    • binary output
    • pulse output
    |• frequency output
    • binary output
    • pulse output
    | |number| |3|3|3|3|3|3| |operating parameters| |5...30 V/< 100 mA|5...30 V/< 100 mA|5...30 V/< 100 mA|5...30 V/< 100 mA|5...30 V/< 100 mA|5...30 V/< 100 mA|

    Page 4

    |frequency output

    • range|kHz|0...5|0...5|0...5|0...5|0...5|0...5| |binary output

    • binary output as alarm output| |limit, change of flow direction or error|limit, change of flow direction or error|limit, change of flow direction or error|limit, change of flow direction or error|limit, change of flow direction or error|limit, change of flow direction or error| |pulse output

    • functions
    • pulse value
    • pulse width
    |units ms|mainly for totalising 0.01...1000 0.05...1000|mainly for totalising 0.01...1000 0.05...1000|mainly for totalising 0.01...1000 0.05...1000|mainly for totalising 0.01...1000 0.05...1000|mainly for totalising 0.01...1000 0.05...1000|mainly for totalising 0.01...1000 0.05...1000| |inputs|inputs|inputs|inputs|inputs|inputs|inputs|inputs| | | |The inputs are galvanically isolated from the transmitter.|The inputs are galvanically isolated from the transmitter.|The inputs are galvanically isolated from the transmitter.|The inputs are galvanically isolated from the transmitter.|The inputs are galvanically isolated from the transmitter.|The inputs are galvanically isolated from the transmitter.| |number| |max. 4, on request|max. 4, on request|max. 4, on request|max. 4, on request|max. 4, on request|max. 4, on request| |• temperature input|• temperature input|• temperature input|• temperature input|• temperature input|• temperature input|• temperature input|• temperature input| |type| |Pt100/Pt1000|Pt100/Pt1000|Pt100/Pt1000|Pt100/Pt1000|Pt100/Pt1000|Pt100/Pt1000| |connection| |4-wire|4-wire|4-wire|4-wire|4-wire|4-wire| |range|°C|-150...+560|-150...+560|-150...+560|-150...+560|-150...+560|-150...+560| |resolution|K|0.01|0.01|0.01|0.01|0.01|0.01| |accuracy| |±0.01 % MV ±0.03 K|±0.01 % MV ±0.03 K|±0.01 % MV ±0.03 K|±0.01 % MV ±0.03 K|±0.01 % MV ±0.03 K|±0.01 % MV ±0.03 K| |• current input|• current input|• current input|• current input|• current input|• current input|• current input|• current input| |accuracy| |0.1 % MV ±10 μA|0.1 % MV ±10 μA|0.1 % MV ±10 μA|0.1 % MV ±10 μA|0.1 % MV ±10 μA|0.1 % MV ±10 μA| |active input

    • range|mA|Uint = 24 V, Rint = 50 Ω, Pint < 0.5 W, not short-circuit proof 0...20|Uint = 24 V, Rint = 50 Ω, Pint < 0.5 W, not short-circuit proof 0...20|Uint = 24 V, Rint = 50 Ω, Pint < 0.5 W, not short-circuit proof 0...20|Uint = 24 V, Rint = 50 Ω, Pint < 0.5 W, not short-circuit proof 0...20|Uint = 24 V, Rint = 50 Ω, Pint < 0.5 W, not short-circuit proof 0...20|Uint = 24 V, Rint = 50 Ω, Pint < 0.5 W, not short-circuit proof 0...20| |passive input

    • range|mA|Rint = 50 Ω, Pint < 0.3 W

    -20...+20|Rint = 50 Ω, Pint < 0.3 W

    -20...+20|Rint = 50 Ω, Pint < 0.3 W

    -20...+20|Rint = 50 Ω, Pint < 0.3 W

    -20...+20|Rint = 50 Ω, Pint < 0.3 W

    -20...+20|Rint = 50 Ω, Pint < 0.3 W

    -20...+20| |• voltage input|• voltage input|• voltage input|• voltage input|• voltage input|• voltage input|• voltage input|• voltage input| |range|V|0...1|0...1|0...1|0...1|0...1|0...1| |accuracy| |0.1 % MV ±1 mV|0.1 % MV ±1 mV|0.1 % MV ±1 mV|0.1 % MV ±1 mV|0.1 % MV ±1 mV|0.1 % MV ±1 mV| |internal resistance| |Rint = 1 MΩ|Rint = 1 MΩ|Rint = 1 MΩ|Rint = 1 MΩ|Rint = 1 MΩ|Rint = 1 MΩ| |• binary input|• binary input|• binary input|• binary input|• binary input|• binary input|• binary input|• binary input| |switching signal| |5...30 V, 1 mA|5...30 V, 1 mA|5...26 V, 1 mA|5...30 V, 1 mA|5...30 V, 1 mA|5...26 V, 1 mA| |functions| |• reset of the measured values
    • reset of the totalisers
    • stop of the totalisers
    • activation of the measuring mode for highly dynamic flows
    |• reset of the measured values
    • reset of the totalisers
    • stop of the totalisers
    • activation of the measuring mode for highly dynamic flows
    |• reset of the measured values
    • reset of the totalisers
    • stop of the totalisers
    • activation of the measuring mode for highly dynamic flows
    |• reset of the measured values
    • reset of the totalisers
    • stop of the totalisers
    • activation of the measuring mode for highly dynamic flows
    |• reset of the measured values
    • reset of the totalisers
    • stop of the totalisers
    • activation of the measuring mode for highly dynamic flows
    |• reset of the measured values
    • reset of the totalisers
    • stop of the totalisers
    • activation of the measuring mode for highly dynamic flows
    |

    ####### 1 with aperture calibration of the transducers 2 for transit time difference principle and reference conditions 3 outside the explosive atmosphere (housing cover open)

    #### Dimensions

  • *72*-***-*A
  • *72*-***-*S
  • 94

    9 x 5ø6.2

    wall mount

    Page 5

    | | | | |---|---|---| | | | |

    255

    | | | |---|---| | | |

    | | | |---|---| |320

    | |

    9 x 5

    190.2

    thread: 6x M20 x 1.5 cable gland: max. 6x M20

    in mm

    93

    165.6

    280 87

    fixing holes for wall mounting

    240

    255

    320

    Page 6

    cable gland: max. 6x M20 with flat gasket and counter nut

    in mm

    2" pipe mounting kit

    Storage

  • • do not store outdoors
  • • store within the original package
  • • store in a dry and dust-free place
  • • protect against sunlight
  • • keep all openings closed
  • |*72*-***-*A

    |item number: 721037-4| |---|---| |*72*-***-*S

    |item number: 721110-4|

  • • storing temperature: -20...+60 °C
  • #### Terminal assignment

    ###### *72*

    AGN

    ARS

    AVS

    AR

    AV

    BGN

    BRS

    Page 7

    BVS

    BR

    BV

    X_AV X_AR X_BV X_BR

    X2 X3

    | | | |---|---|

    | | | |---|---|

    LAN USB

    T1A

    T1B

    T2A

    T2B

    T3A

    T3B

    T4B T4A

    S1

    S3

    Page 8

    | | | | | | | | | | | | |---|---|---|---|---|---|---|---|---|---|---| |T1a|T1b|S1|T2a|T2b|T3a|T3b|S3|T4a|T4b| |

    P1+

    P2+

    P3+

    P4+

    P6a

    P7a

    P5a

    A+

    B-

    P5b

    P6b

    P7b

    P1-

    P2-

    P3-

    P4-

    S

    Page 9

    S

    | | | | | | | |---|---|---|---|---|---| |PE| |N(-)| |L(+)| |

    | | | | |---|---|---| | | | |

    *72*-*-*S *72*-*-*A equipotential bonding terminal

    |power supply1|power supply1|power supply1|power supply1|power supply1|power supply1|power supply1|power supply1|power supply1|power supply1|power supply1|power supply1|power supply1|power supply1|power supply1|power supply1|power supply1|power supply1|power supply1| |---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---| |terminal|terminal|terminal|terminal|terminal|connection (AC)|connection (AC)|connection (AC)|connection (AC)|connection (AC)|connection (AC)|connection (AC)|connection (AC)|connection (DC)|connection (DC)|connection (DC)|connection (DC)|connection (DC)|connection (DC)| |PE|PE|PE|PE|PE|protective conductor|protective conductor|protective conductor|protective conductor|protective conductor|protective conductor|protective conductor|protective conductor|protective conductor|protective conductor|protective conductor|protective conductor|protective conductor|protective conductor| |N(-)|N(-)|N(-)|N(-)|N(-)|neutral conductor|neutral conductor|neutral conductor|neutral conductor|neutral conductor|neutral conductor|neutral conductor|neutral conductor|-|-|-|-|-|-| |L(+)|L(+)|L(+)|L(+)|L(+)|outer conductor|outer conductor|outer conductor|outer conductor|outer conductor|outer conductor|outer conductor|outer conductor|+|+|+|+|+|+| |transducers|transducers|transducers|transducers|transducers|transducers|transducers|transducers|transducers|transducers|transducers|transducers|transducers|transducers|transducers|transducers|transducers|transducers|transducers| |transducer cable (transducers *8*, **LI*), extension cable|transducer cable (transducers *8*, LI*), extension cable|transducer cable (transducers *8*, LI*), extension cable|transducer cable (transducers *8*, LI*), extension cable|transducer cable (transducers *8*, LI*), extension cable|transducer cable (transducers *8*, LI*), extension cable|transducer cable (transducers *8*, LI*), extension cable|transducer cable (transducers *8*, LI*), extension cable|transducer cable (transducers *8*, LI*), extension cable| | | |transducer cable (transducers *52)|transducer cable (transducers *52)|transducer cable (transducers *52)|transducer cable (transducers *52)|transducer cable (transducers *52)|transducer cable (transducers *52)|transducer cable (transducers ***52)| |measuring channel A|measuring channel A|measuring channel A|measuring channel A|measuring channel B|measuring channel B|measuring channel B|measuring channel B|measuring channel B| | | |measuring channel A|measuring channel A|measuring channel A|measuring channel B|measuring channel B|measuring channel B| | |terminal|connection|connection|connection|terminal|terminal|connection|connection|connection|transducer|transducer|transducer|terminal|terminal|terminal|terminal|terminal|terminal|connection| |AV|signal|signal|signal|BV|BV|signal|signal|signal|![image 7](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile7.png)|![image 7](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile7.png)|![image 7](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile7.png)|X_AV|X_AV|X_AV|X_BV|X_BV|X_BV|SMB connector| |AVS|shield|shield|shield|BVS|BVS|shield|shield|shield|![image 7](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile7.png)|![image 7](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile7.png)|![image 7](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile7.png)|X_AV|X_AV|X_AV|X_BV|X_BV|X_BV|SMB connector| |ARS|shield|shield|shield|BRS|BRS|shield|shield|shield|![image 8](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile8.png)|![image 8](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile8.png)|![image 8](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile8.png)|X_AR|X_AR|X_AR|X_BR|X_BR|X_BR|SMB connector| |AR|signal|signal|signal|BR|BR|signal|signal|signal|![image 8](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile8.png)|![image 8](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile8.png)|![image 8](product-data-sheet-flexim-fluxus-f722-non-intrusive-synchronized-channel-averaging-sca-flow-meter-en-10177690_images/imageFile8.png)|X_AR|X_AR|X_AR|X_BR|X_BR|X_BR|SMB connector| |outputs1, 2|outputs1, 2|outputs1, 2|outputs1, 2|outputs1, 2|outputs1, 2|outputs1, 2|outputs1, 2|outputs1, 2|outputs1, 2|outputs1, 2|outputs1, 2|outputs1, 2|outputs1, 2|outputs1, 2|outputs1, 2|outputs1, 2|outputs1, 2|outputs1, 2| |terminal|terminal|connection|connection|connection|connection|connection|connection| | |terminal|terminal|terminal|terminal|connection|connection|connection|communication interface

    1|communication interface

    1| |P1+...P4+ P1-...P4-|P1+...P4+ P1-...P4-|current output, voltage output, frequency output, HART (P1)|current output, voltage output, frequency output, HART (P1)|current output, voltage output, frequency output, HART (P1)|current output, voltage output, frequency output, HART (P1)|current output, voltage output, frequency output, HART (P1)|current output, voltage output, frequency output, HART (P1)| | |A+|A+|A+|A+|signal +|signal +|signal +|• RS485
    • Modbus RTU1
    • BACnet MS/TP1
    • M-Bus1
    • Profibus PA1
    • FF H11
    |• RS485
    • Modbus RTU1
    • BACnet MS/TP1
    • M-Bus1
    • Profibus PA1
    • FF H11
    | |P1+...P4+ P1-...P4-|P1+...P4+ P1-...P4-|current output, voltage output, frequency output, HART (P1)|current output, voltage output, frequency output, HART (P1)|current output, voltage output, frequency output, HART (P1)|current output, voltage output, frequency output, HART (P1)|current output, voltage output, frequency output, HART (P1)|current output, voltage output, frequency output, HART (P1)| | |B-|B-|B-|B-|signal -|signal -|signal -|• RS485
    • Modbus RTU1
    • BACnet MS/TP1
    • M-Bus1
    • Profibus PA1
    • FF H11
    |• RS485
    • Modbus RTU1
    • BACnet MS/TP1
    • M-Bus1
    • Profibus PA1
    • FF H11
    | |P5a...P7a
    P5b...P7b
    |P5a...P7a
    P5b...P7b
    |digital output|digital output|digital output|digital output|digital output|digital output| | |S|S|S|S|shield|shield|shield|• RS485
    • Modbus RTU1
    • BACnet MS/TP1
    • M-Bus1
    • Profibus PA1
    • FF H11
    |• RS485
    • Modbus RTU1
    • BACnet MS/TP1
    • M-Bus1
    • Profibus PA1
    • FF H11
    | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |USB|USB|USB|USB|type B Hi-Speed USB 2.0 Device|type B Hi-Speed USB 2.0 Device|type B Hi-Speed USB 2.0 Device|• service (FluxDiag/ FluxDiagReader)|• service (FluxDiag/ FluxDiagReader)| | | | | | | | | | | |LAN|LAN|LAN|LAN|RJ45 10/100 Mbps Ethernet|RJ45 10/100 Mbps Ethernet|RJ45 10/100 Mbps Ethernet|• service (FluxDiag/ FluxDiagReader)
    • BACnet IP
    • Modbus TCP
    |• service (FluxDiag/ FluxDiagReader)
    • BACnet IP
    • Modbus TCP
    | |analog inputs1, 2|analog inputs1, 2|analog inputs1, 2|analog inputs1, 2|analog inputs1, 2|analog inputs1, 2|analog inputs1, 2|analog inputs1, 2|analog inputs1, 2|analog inputs1, 2|analog inputs1, 2|analog inputs1, 2|analog inputs1, 2|analog inputs1, 2|analog inputs1, 2|analog inputs1, 2|analog inputs1, 2|analog inputs1, 2|analog inputs1, 2| | | | |temperature probe|temperature probe|temperature probe|temperature probe|temperature probe|temperature probe|temperature probe|temperature probe|passive sensor|passive sensor|passive sensor|passive sensor|passive sensor|active sensor|active sensor|active sensor| |terminal|terminal|terminal|direct connection|direct connection|direct connection|direct connection|connection with extension cable|connection with extension cable|connection with extension cable|connection with extension cable|connection|connection|connection|connection|connection|connection|connection|connection|

    Page 10

    |T1a...T4a|T1a...T4a|T1a...T4a|red|red|red|red|red|red|red|red|not connected|not connected|not connected|not connected|not connected|not connected|not connected|not connected| |T1A...T4A|T1A...T4A|T1A...T4A|red/blue|red/blue|red/blue|red/blue|grey|grey|grey|grey|-|-|-|-|-|+|+|+| |T1b...T4b|T1b...T4b|T1b...T4b|white/blue|white/blue|white/blue|white/blue|blue|blue|blue|blue|+|+|+|+|+|not connected|not connected|not connected| |T1B...T4B|T1B...T4B|T1B...T4B|white|white|white|white|white|white|white|white|not connected|not connected|not connected|not connected|not connected|-|-|-| |S1, S3|S1, S3|S1, S3|shield|shield|shield|shield|shield|shield|shield|shield|not connected|not connected|not connected|not connected|not connected|not connected|not connected|not connected| |binary inputs1, 2|binary inputs1, 2|binary inputs1, 2|binary inputs1, 2|binary inputs1, 2|binary inputs1, 2|binary inputs1, 2|binary inputs1, 2|binary inputs1, 2|binary inputs1, 2|binary inputs1, 2|binary inputs1, 2|binary inputs1, 2|binary inputs1, 2|binary inputs1, 2|binary inputs1, 2|binary inputs1, 2|binary inputs1, 2|binary inputs1, 2| |terminal|terminal|terminal|terminal|terminal|terminal|terminal|terminal|terminal|terminal|terminal|terminal|terminal|terminal|terminal|terminal|terminal|terminal|terminal| |P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|P1+...P2+, P1-...P2-|

  • 1 сable (by customer):
  • - e.g. flexible wires, with insulated wire ferrules, wire cross-section: 0.25...2.5 mm2
  • - outer diameter of the cable (*72*-***-*S with ferrite nut): max. 7.6 mm
  • 2 The number, type and terminal assignment are customised.
  • Transducers Overview Shear wave transducers

    | | | |technical type|technical type|technical type|technical type|technical type|technical type| |---|---|---|---|---|---|---|---|---| | | | |G|K|M|P|Q|S| | | | | | | | | | | |zone 2 - FM Class I Div. 2 - nonEx normal temperature range|zone 2 - FM Class I Div. 2 - nonEx normal temperature range|zone 2 - FM Class I Div. 2 - nonEx normal temperature range|CDG1N52 CLG1N52|CDK1N52 CLK1N52|CDM2N52 CLM2N52|CDP2N52 CLP2N52|CDQ2N52 CLQ2N52|CDS2N52| | | | | | | | | | | |zone 2 - nonEx IP68|zone 2 - nonEx IP68|zone 2 - nonEx IP68|CDG1LI8|CDK1LI8|CDM2LI8|CDP2LI8| | | | | | | | | | | | | |zone 2 - FM Class I Div. 2 - nonEx extended temperature range|zone 2 - FM Class I Div. 2 - nonEx extended temperature range|zone 2 - FM Class I Div. 2 - nonEx extended temperature range|CDG1E52 CLG1E52|CDK1E52 CLK1E52|CDM2E52 CLM2E52|CDP2E52 CLP2E52|CDQ2E52 CLQ2E52| | | | | | | | | | | | |zone 1 normal temperature range|zone 1 normal temperature range|zone 1 normal temperature range|CDG1N81 CLG1N81|CDK1N81 CLK1N81|CDM2N81 CLM2N81|CDP2N81 CLP2N81|CDQ2N81 CLQ2N81| | | | | | | | | | | | |zone 1 IP68|zone 1 IP68|zone 1 IP68|CDG1LI1|CDK1LI1|CDM2LI1|CDP2LI1| | | | | | | | | | | | | |zone 1 extended temperature range|zone 1 extended temperature range|zone 1 extended temperature range|CDG1E83 CLG1E83|CDK1E83 CLK1E83|CDM2E85 CLM2E85|CDP2E85 CLP2E85|CDQ2E85 CLQ2E85| | | | | | | | | | | | |inner pipe diameter d|inner pipe diameter d|inner pipe diameter d|inner pipe diameter d|inner pipe diameter d|inner pipe diameter d|inner pipe diameter d|inner pipe diameter d| | |min. extended|mm| |400|100|50|25|10|6|

    Page 11

    |min. recommended|mm| |500|200|100|50|25|10| |max. recommended|mm| |4000|2000|1000|400|150|70| |max. extended|mm| |6500|2400|1200|480|240|70| |pipe wall thickness|pipe wall thickness|pipe wall thickness|pipe wall thickness|pipe wall thickness|pipe wall thickness|pipe wall thickness|pipe wall thickness| | |min.|mm| |11|5|2.5|1.2|0.6|0.3|

    for further data see Technical specification TS_F7xx-transducersVx-xXX_Leu

    #### Transducer mounting fixture

    |Variofix L

    |Variofix L

    |Variofix C|transducer box WI for WaveInjector with chains| |---|---|---|---| | |transducer frequency S

    | | | | | | | | | | |Variofix C with bolt mounting plates|transducer box WI for WaveInjector with threaded rods| | | |outer pipe diameter:

    VCM: max. 46 mm VCQ: max. 36 mm

    |outer pipe diameter: 35...380 mm

    |

    for further data see Technical specification TS_F7xx-transducersVx-xXX_Leu

    #### Coupling materials for transducers

    | |normal temperature range|normal temperature range|extended temperature range|extended temperature range|extended temperature range|WaveInjector|WaveInjector| |---|---|---|---|---|---|---|---| | |< 100 °C|< 170 °C|< 150 °C|< 200 °C|200...240 °C|< 280 °C|280...630 °C| |< 24 h|coupling compound type N or coupling foil type VT|coupling compound type E or coupling foil type VT|coupling compound type E or coupling foil type VT|coupling compound type E or H or coupling foil type VT|coupling foil type TF|coupling foil type A and coupling foil type VT|coupling foil type B and coupling foil type VT| |long time measurement|coupling foil type VT|coupling foil type VT|coupling foil type VT|coupling foil type VT|coupling foil type TF|coupling foil type A and coupling foil type VT|coupling foil type B and coupling foil type VT|

    for further data see Technical specification TS_F7xx-transducersVx-xXX_Leu

    #### Connection systems

    |connection system TS|connection system TS|connection system TS| |---|---|---| |connection with extension cable|direct connection|transducers technical type| |transmitter

    JB02, JB03, JB04

    x

    l|transmitter

    x|*52| |connection system T1|connection system T1|connection system T1|

    Page 12

    |connection with extension cable|direct connection|transducers technical type| |transmitter

    JB01 x

    l|transmitter

    x|*8*| |transmitter

    x

    l

    JB01, JBP2, JBP3|transmitter

    x

    |****LI*|

    for further data see Technical specification TS_F7xx-transducersVx-xXX_Leu

    Temperature probes

    |PT12N|PT12N|PT12F| |---|---|---| |item number:

    • 770415-1
    • 770414-2 (matched)
    |item number:

    • 770415-1A2
    • 770414-1A2 (matched)
    |item number:

    • 770415-2| |• Pt100
    • clamp-on
    • -30...+250 °C
    |• Pt100
    • clamp-on
    • -30...+250 °C
    • ATEX
    |• Pt100
    • clamp-on
    • -45...+250 °C
    • response time: 8 s
    | |direct connection

    |direct connection

    |direct connection

    | |connection with extension cable

    extension cable

    junction box|connection with extension cable

    extension cable

    junction box|connection with extension cable

    extension cable

    junction box|

    see Technical specification TS_PTVx-xXX_Leu

    Annex Reference conditions as available at e.g. the test facilities of Physikalisch-Technische Bundesanstalt

    |measurement principle| |transit time difference correlation principle| |---|---|---| |all uncertainties|%|95| |fluid temperature| |25 °C ±5 K| |ambient temperature| |25 °C ±5 K| |warm-up time|min|10| |flow profile at the measuring point| |fully developed, rotationally symmetric| |installation| |installation according to specifications using the recommended transducers| |Reynolds number| |> 10 000| |pipe diameter uncertainty|%|0.2| |pipe wall thickness uncertainty|%|1| |circularity tolerance| |0.08 % of inner pipe diameter| |SCNR|dB|> 48| |SNR|dB|> 12|

    Technical Specification TSFLUXUS_F72xV3-0-0-0EN_Leu

    For more information: Emerson.com © 2023 Emerson. All rights reserved. Emerson Terms and Conditions of Sale are available upon request. The Emerson logo is a trademark and service mark of Emerson Electric Co. Flexim is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners.