warning Labcold

Error Code 3 blinks - low battery alarm

Red LED blinks three times indicating low battery charge level for controller.

? What Causes This

Battery depletion to a critical but not fully dead state, usually due to discharge over long periods without recharge or inadequate initial charge at installation.

How to Fix It

Replace with new batteries (4 x AA 1.5V alkaline) as soon as possible, preferably even before the alarm sounds for prolonged uptime safety and prevent potential controller loss of function during outages.

Source: page 26 of the official manual

From the manual

Manual page 26 — Labcold error 3 blinks - low battery alarm

This error code appears in: