error TD Gluco
Error Code E-F
Test stopped due to insufficient blood volume or improper removal before completion of the test cycle.
? What Causes This
Inadequate sample size applied, premature removal of strip during countdown or incorrect handling when applying sample.
✓ How to Fix It
Perform test using a larger volume sample; ensure sample application only occurs when appropriate signal appears on screen. If issue continues, contact GlucoRx Customer care on 01483 755133.
Source: page 29 of the official manual
From the manual
This error code appears in:
TD Gluco TD 4279 Blood Sugar Control System