warning Visage
Error Code The cuff is a little bit low, please heighten the cuff.
Voice command from device if wristband placed too low on patient’s arm.
? What Causes This
Improper positioning of the blood pressure cuff below a good part of the upper arm circumference area or misplacement to cover chest instead of arm.
✓ How to Fix It
Re-position and correctly fit the blood pressure cuff again, as detailed in manual page 35.
Source: page 44 of the official manual
From the manual
This error code appears in:
Visage MD 15469