What is the 'd-o' alarm and how do I clear it?

The 'd-o' alarm is displayed if the door remains open for more than 90 seconds, at which point the audible alarm will sound. Unlike some other alarms, this one does not cancel automatically when the door is closed — you must press the Mute button to clear it and turn off the red alarm indicator. (Page 18–19)

From the official Labcold Intellicold manual, page 18

Have a follow-up question?

Answers from the official manual.